For example: Agilent E4440A or Anritsu 68369
Hipots

Model:HT7000DC-200mA

Manufactured by: Compliance West

Category:Ground Continuity Testers / Hi-Pot

Compliance West HT7000DC-200mA - Ground Continuity Testers / Hi-Pot

Compliance West HT7000DC-200mA Overview

Our HT7000DC-200mA has DC output only. Leakage current limits can be set with the rear panel knob from 10mA to 200mA. Normal operation stops testing when a leakage current greater than the set amount or when a dielectric breakdown is encountered. If desired, testing can be set to continue to burn problem areas. Testing will continue until the design current limit of the tester is exceeded; at which time the output voltage will return to zero.
Packaged in a 6U enclosure., with all controls and switches on the front panel for easier access. Provided with test leads terminated in alligator clips and a calibration certificate.

In any device, lower ripple voltages require larger filter capacitors, which store more power, and that power is normally used to smooth the rectified waveform. In the case of a hipot tester, this stored power has the side effect of discharging into the EUT in the event of a failure. In order to keep the power that must be absorbed by the EUT within reasonable limits, we have used three phase input power for this model, which improves ripple without the side effect of excessive power sent to the EUT when a dielectric breakdown is detected.

How can we help you with today?

RECENTLY VIEWED