Model:P6042

Manufactured by: Tektronix

Category:Probes

Tektronix P6042 - Probes

Tektronix P6042 Overview

FEATURES/CHARACTERISTICS:

  • Bandwidth DC to 50 MHz
  • Aberrations <3% from 1mA/DIV to 50mA/DIV
  • Noise Periodic <0.5mA +0.2 div, tangentially measured
  • Random Trace Shift <1.5mA
  • Dynamic Range +500mV to -500mV at probe amplifier
  • Positioning Range +300mV to -300mV at probe amplifier output
  • Attenuator Accuracy within +/-3%
  • Required Output Load Impedance 50 Ohm
  • Input Power <20W
  • Risetime 7ns
  • Maximum Input Current 1A to 5mA: 10A (DC+peak AC), depending on frequency. 2mA and 1mA: 0.5A (DC+peak AC)
  • Thermal Drift (Probe only) <2mA/°C from 15°C to 35°C with amplifier temperature constant.
  • Thermal Drift (Amplifier only) <0.2div/°C with probe temperature constant.
  • Voltage-Feedthrough Susceptibility <250µA/V at 50MHz
  • External Magnetic Field Suscceptibility <250µA/Gauss, at 60Hz A very important Instrument in every laboratory, a current probe to measure from DC to higher frequencies. The Tektronix P6042 was designed for use with scopes having either 50 Ohm or high-impedance inputs. Set the scope to 50mV/DIV, terminate the coaxial cable with 50 Ohm. For somebody developing a switching power supply, a current probe is a big help.